Example Capillary Measurement SOP¶
- Plug the CapiScope HVCS USB plug into the laptop.
- Click on the CapiScope HVCS desktop icon.
- For the CapiScope Select Profile box click OK
- Click the record button (or f12 key). Click again to stop.
- Open the Image sequence dialog box.
- Enter a new base filename and path for recordings. (Use the browse button at right to browse and create new folders).
- Click OK
- Press the start button to start the camera.
- Focus using focus ring (sublingual) or top positioner (nailfold) and adjust brightness.
- Add oil if viewing nailfold.
- Repeat 9 and locate measuring site.
- Press record button to start recording loop.
- Watch time elapsed as shown in bottom line of the status box.
- Press the pause button to stop recording.
- Goto 8 to start next recording with the same base filename.
Analysing Video¶
- Open the video file.
- Press the lower play button, or drag the horizontal scrollbar to the first good image in focus.
- Select menu edit, duplicate image in new window.
- Press measurement button to show the dimension line buttons.
- Press the dimension line button.
- Draw vessel diameter measuring lines across apex of each capillary in final row.
- Press TAB key until first dimension line is highlighted.
- Press the find width button.
- Press accept ( or return key )if the greylevel window shows a clean “V” and the sides of the V have been detected ok.
- Repeat 7-9 for the remaining capillaries.
- Save the Image with the dimension lines.
- Save the Diameter Calculation Results.
- Go back to the video file.
- Use Tools, Automatic Movement Correction.
- Press <ctrl> l to play the recording in a continuous loop.
- Draw a dimension line on the first capillary in the final row. Draw the line in the direction of flow.
- Press the stop button.
- Use Window, Linescan to open the linescan window.
- Activate the video file window and press play button.
- This will produce a velocity trace in the linescan window as a black line.
- Select a portion or all the trace for averaging by dragging while holding down the left mouse button.
- Use the TAB key to step through the different traces in the linescan window until the status box top line shows 0:cbv. (capilllary blood cell velocity)
- Press = button or = key to get the average cbv.
- Activate the video file window and press TAB to activate the dimension line. Press delete to delete it.
- Repeat steps 15-17, 19-24 for each other capillary to measure in the final row.
- Save the chart results window with the cbv averages.